Recall of QUICKDRAW 25 VENOUS CANNULA

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by EDWARDS LIFESCIENCES (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74042
  • Event Risk Class
    II
  • Event Initiated Date
    2013-06-07
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Risks associated with specific removal techniques: (1) scar tissue at the incision site may create resistance and increase the withdrawal forces on the cannula. (2) when using the percutaneous technique separation of the cannula may occur during removal if too much compressive force is applied at the incision site. this is in addition to the resistance felt if scar tissue is present.

Device

  • Model / Serial
    Model Catalog: QD25 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: QD22 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    QUICKDRAW VENOUS CANNULA MODEL QD25
  • Manufacturer

Manufacturer