Recall of PRESSURE REGULATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by JOHNSON & JOHNSON MEDICAL PRODUCTS A DIV. OF JOHNSON & JOHNSON INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27354
  • Event Risk Class
    II
  • Event Initiated Date
    2016-03-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The current recommendation of spraying evicel (fibrin sealant) with the 45cm tip during laparoscopy surgery is at a pressure range of 1.4 to 1.7bar. the european medicines agency have clarified their recommendation for pressure to be used when spraying evicel with the 45cm tip during laparoscopic surgery from a range of 1.4-1.7 bar to 1.4 bar only. a similar approach has been taken in canada. thus the labelling of the pressure regulators will be updated and the sticker label on each pressure regulator will be replaced to reflect this change.

Device

Manufacturer