Recall of PRECISION 500D SYSTEM - FLURO CARRAGE & TOWER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    27344
  • Event Risk Class
    III
  • Event Initiated Date
    2013-12-24
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    It has been identified during several preventive maintenance activities that the 4 outer screws that hold the overhead video monitor suspension to the carriage were found missing since installation of the system. there is a potential that your system has not been installed per specified installation requirements. ge healthcare is not aware of any fall of the overhead monitor suspension. however missing screws may increase the risk for fall of the monitor suspension. no injury has been reported due to this issue. missing screws may increase the risk for the monitor to fall.

Device

  • Model / Serial
    Model Catalog: S0910ZC (Lot serial: n/a); Model Catalog: S0912PB (Lot serial: n/a); Model Catalog: S14501BT (Lot serial: n/a); Model Catalog: S13001AX (Lot serial: n/a); Model Catalog: 2258627 (Lot serial: n/a)
  • Product Description
    PRECISION 500D SYSTEM
  • Manufacturer

Manufacturer