Recall of POWER PRO IT AMBULANCE COT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    105570
  • Event Risk Class
    II
  • Event Initiated Date
    2016-06-23
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A few service reports alleged that "smoke" was coming from the foot end of the cot. this reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. in addition to the alleged smoking one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later it began to smoke.

Device

  • Model / Serial
    Model Catalog: 6516000000 (Lot serial: 150840805); Model Catalog: 6500002014 (Lot serial: N/A); Model Catalog: 6506000000 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    POWER PRO IT AMBULANCE COT;CONTROL BOARD FOR POWER PRO AMBULANCE COT;POWER PRO AMBULANCE COT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    HAMILTON
  • Manufacturer Parent Company (2017)
  • Source
    HC