Recall of PLUM 360 INFUSION SYSTEM WITH MEDNET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOSPIRA HEALTHCARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    109052
  • Event Risk Class
    I
  • Event Initiated Date
    2017-01-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
  • Reason
    When the ac power is lost and quickly resumed the pump software incorrectly classifies the event as an e323 alarm condition resulting in an interruption of any ongoing therapy. if this issue occurs the pump requires manual user input to resume previous infusion.