Recall of PERIACETABULAR RECONSTRUCTION PROSTHESIS (PAR) - CUSTOM MADE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOWMEDICA OSTEONICS CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    133417
  • Event Risk Class
    II
  • Event Initiated Date
    2013-09-16
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    This is a sap recall not web posted. insufficient taper torsional strength may result when a par extension piece is used with a forged cocr stem. stryker received two reports indicating rotation or disassociation of a par extension piece when used in combination with a forged cocr stem. in each case rotation or dissociation was reported to occur upon initial weight bearing post-surgery.

Device

  • Model / Serial
    Model Catalog: C-M071-4-503 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M071-8-001 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M071-8-803 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M072-0-901 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M072-0-902 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M072-0-903 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M078-6-002 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M079-0-602 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M071-4-502 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M071-4-501 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M069-2-802 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M068-2-403 (Lot serial: >10 numbers contact mfg); Model Catalog: C-M068-2-402 (Lot serial: >10 numbers contact mfg); Model Catalog: C-KM40-1-303 (Lot serial: >10 numbers contact mfg); Model Catalog: C-KM40-1-302 (Lot serial: >10 numbers contact mfg); Model Catalog: C-K
  • Product Description
    PAR Extension Piece - Custom Implant SAP
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MAHWAH
  • Manufacturer Parent Company (2017)
  • Source
    HC