Recall of PARADIGM QUICK-SET INFUSION SETS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by INSULINPUMPS.CA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    54531
  • Event Risk Class
    I
  • Event Initiated Date
    2009-07-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Approximately 2% of infusion sets in the affected lots (lots starting with '8') may not allow the insulin pump to vent properly. venting is necessary to equalize the pressure in the reservoir compartment with the surrounding atmosphere. if the vent does not work properly this could result in too much or too little insulin being delivered.

Device

  • Model / Serial
    Model Catalog: MMT-397 (Lot serial: CONTACT MFR); Model Catalog: MMT-398 (Lot serial: CONTACT MFR); Model Catalog: MMT-396 (Lot serial: CONTACT MFR); Model Catalog: MMT-399 (Lot serial: CONTACT MFR)
  • Product Description
    PARADIGM QUICK-SET INFUSION SETS
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Source
    HC