Recall of PAK ULTRASOUND FLUIDICS MANAGEMENT SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ALCON CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    80902
  • Event Risk Class
    I
  • Event Initiated Date
    2014-11-04
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Alcon received complaints for the infiniti ultrasound paks related to occlusion of the aspiration line. the quality review processes showed an equipment misalignment which may have potentially detached a piece of the aspiration tubing and inserted it into the aspiration line of the infiniti ultrasound paks. this recall is being conducting due to the fact that the possibility exists for a loss of aspiration performance during surgery which may pose a safety risk to patients.

Device

  • Model / Serial
    Model Catalog: 8065-7410-80 (Lot serial: 1642478H); Model Catalog: 8065-7410-81 (Lot serial: 1623756H); Model Catalog: 8065-7410-83 (Lot serial: 1623756H); Model Catalog: 8065-7410-83 (Lot serial: 1647970H); Model Catalog: 8065-7410-81 (Lot serial: 1647970H); Model Catalog: 8065-7410-80 (Lot serial: 1647970H); Model Catalog: 8065-7410-83 (Lot serial: 1642478H); Model Catalog: 8065-7410-81 (Lot serial: 1642478H); Model Catalog: 8065-7410-80 (Lot serial: 1623756H); Model Catalog: 8065-7410-83 (Lot serial: 1642433H); Model Catalog: 8065-7410-81 (Lot serial: 1642433H); Model Catalog: 8065-7410-80 (Lot serial: 1642433H); Model Catalog: 8065-7410-83 (Lot serial: 1647977H); Model Catalog: 8065-7410-81 (Lot serial: 1647977H); Model Catalog: 8065-7410-80 (Lot serial: 1647977H)
  • Product Description
    PAK ULTRASOUND FLUIDICS MANAGEMENT SYSTEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC