Recall of OXYARM - OXYMASK MODEL -RTM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SOUTHMEDIC INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28832
  • Event Risk Class
    II
  • Event Initiated Date
    2017-02-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    On february 14 southmedic received a customer complaint indicating a potential occlusion of the co2 sampling line on a southmedic capnography mask. inspection of current stock of the lot number in question did not identify any issues however investigation into additional lots manufactured with the same subassembly resulted in identification of occluded co2 sampling lines. subsequently all stock of various product skus manufactured with a co2 sampling line were placed on hold for further inspection.

Device

  • Model / Serial
    Model Catalog: OM-2125-8 (Lot serial: W35266); Model Catalog: OP-2125-8 (Lot serial: W35909); Model Catalog: CO2-01-SLM (Lot serial: W35983); Model Catalog: CO2-03-SLM (Lot serial: W35983); Model Catalog: CO2-01-SLM (Lot serial: W35906); Model Catalog: CO2-03-SLM (Lot serial: W35906)
  • Product Description
    OxyArm-OxyMask Model - RTM;OxyArm-OxyPlus;Ophthalmic Capnoxygen Mask
  • Manufacturer

Manufacturer