Recall of OPTIONAL 110V OUTLET BOX ASSEMBLY FOR SECURE II HOSPITAL BED

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21180
  • Event Risk Class
    II
  • Event Initiated Date
    2002-05-15
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Nylon screws on the assembly of the 110v outlet option may become loose or break allowing the receptacle to move inside the outlet box and has the potential to create an electrical shock.

Device

  • Model / Serial
    Model Catalog: 3002 (Lot serial: S/NS 10615815-10615816); Model Catalog: 3002 (Lot serial: 11018392 11115346.); Model Catalog: 3002 (Lot serial: 10815431-10815432 10815534); Model Catalog: 3002 (Lot serial: 10715049-10715093)
  • Product Description
    MED/SURG BED OPTIONAL 110V OUT BOS ASSE