Recall of OPTIMA XR220AMX X-RAY SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    26856
  • Event Risk Class
    III
  • Event Initiated Date
    2012-09-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Software defect where under certain instances and protocols the image from the scan will not appear requiring a rescan. the three problematic instances are the following: the customer selects the next protocol in a multiple exposure exam while the previous image is displayed. the detector management window is opened while the protocol is opening on the user interface. the auto protocol assist feature is utilized and one of the protocol codes is not matched to one in the system database.

Device

Manufacturer