Recall of OPTIMA MR450W MRI SYSTEM - MAIN UNIT - 32CH

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    85368
  • Event Risk Class
    II
  • Event Initiated Date
    2013-08-06
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The yellow foam strip used between the end bell covers and patient bore of certain ge mr products may extend past the surface of the covers. the foam itself is not hazardous. however any fluid or body substance absorbed by the extended foam may come into contact with patients and act as a fomite potentially transferring infectious agents from one patient to another.

Device

  • Model / Serial
    Model Catalog: M7000WC (Lot serial: N/A); Model Catalog: M7000BH (Lot serial: N/A); Model Catalog: M7000WB (Lot serial: N/A); Model Catalog: M7000WA (Lot serial: N/A); Model Catalog: M7000BB (Lot serial: N/A); Model Catalog: M7000ZJ (Lot serial: N/A)
  • Product Description
    OPTIMA MR450W MRI SYSTEM -MAIN-32CH
  • Manufacturer

Manufacturer