Recall of OPMI LUMERA 700

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CARL ZEISS CANADA LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    37228
  • Event Risk Class
    III
  • Event Initiated Date
    2012-02-23
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Jamming of cables in the suspension arm may lead to damage of video and/or related cables caused by an abrasion of the cables located in the suspension arm. there is no risk of harm or safety related to this issue. the opmi lumera 700 can be used in manual mode in the worst case scenario.

Device

  • Model / Serial
    Model Catalog: 305953-9920-000 (Lot serial: >10 numbers contact mfg.); Model Catalog: 305953-9900-000 (Lot serial: >10 numbers contact mfg.)
  • Product Description
    OPMI LUMERA 700 Ophthalmic Surgical Microscope
  • Manufacturer

Manufacturer