Recall of OPERA LIFT & SLING

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ARJO LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    25055
  • Event Risk Class
    II
  • Event Initiated Date
    2002-08-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Spirol pin that locks a castle nut to the pivot pin had migrated out allowing the castle nut to unscrew and detach the spreader bar from the jib.

Device

  • Model / Serial
    Model Catalog: (Lot serial: WEEKS PRECEDING 20/02); Model Catalog: (Lot serial: ///); Model Catalog: (Lot serial: #S PRECEEDING GB2002896045002)
  • Product Description
    900353 TEMPO PATIENT LIFT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GLOUCESTER
  • Source
    HC