Recall of ONLINE DAT AMPHETAMINES II FOR COBAS INTEGRA

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ROCHE DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    93145
  • Event Risk Class
    III
  • Event Initiated Date
    2012-06-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Patient and control sample recovery may be decreased with the online dat amphetamines ii reagent (lot number 654236-01) when used on the cobas integra 400 and 800 series analysers. the affected lot shows reduced d-amphetamine cross-reactivity for higner concentrated d-amphetamine samples.

Device

  • Model / Serial
    Model Catalog: 04512936190 (Lot serial: 65423601)
  • Product Description
    ONLINE DAT AMPHETAMINES II FOR COBAS INTEGRA
  • Manufacturer

Manufacturer

  • Manufacturer Address
    LAVAL
  • Manufacturer Parent Company (2017)
  • Source
    HC