Recall of ONETOUCH VERIO IQ BLOOD GLUCOSE METER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIFESCAN EUROPE A DIVISION OF CILAG GMBH INTERNATIONAL.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    104148
  • Event Risk Class
    II
  • Event Initiated Date
    2012-02-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Software issue has been identified under very specific circumstances produces an error that causes the meter to turn itself off when a user attempts to access the "results log" to view stored test results. the meter enters the set-up mode and requires the user to confirm the date and time settings before testing.

Device

Manufacturer