Recall of ONCOR EXPRESSION DIGITAL LINEAR ACCELERATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    25574
  • Event Risk Class
    II
  • Event Initiated Date
    2017-01-31
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    It has come to our attention that for digital linear accelerators of type artiste oncor and primus with serial numbers equal or higher than 3094: 1) incorrect values for the rotational collimator position may be displayed due to mechanical problems of the sensor slide 2) the overhead suspension might break due to an improper weld seam which might result in an injury of the patient.

Device

  • Model / Serial
    Model Catalog: 7360717 (Lot serial: SN = OR > 3094); Model Catalog: 8139789 (Lot serial: SN = OR > 3094); Model Catalog: 04504200 (Lot serial: SN = OR > 3094)
  • Product Description
    ONCOR EXPRESSION DIGITAL LINEAR ACCELERATOR;ARTISTE MV-LINEAR ACCELERATOR-MAIN UNIT;MEVATRON LINEAR ACCELERATORS
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source