Recall of ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by NUCLETRON B.V. (A WHOLLY OWNED SUBSIDIARY OF ELEKTA AB).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74401
  • Event Risk Class
    II
  • Event Initiated Date
    2015-12-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    When using the option "tumor overlap fraction" in vmat planning it has been observed that in rare cases the system does include an organ at risk as target volume. this could result in open mlc and open jaws in areas away from the target volume.

Device

Manufacturer

  • Manufacturer Address
    VEENENDAAL
  • Manufacturer Parent Company (2017)
  • Source
    HC