Recall of ON-BOARD IMAGER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by VARIAN MEDICAL SYSTEMS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17189
  • Event Risk Class
    II
  • Event Initiated Date
    2008-10-01
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Sw errors between impac multi-access or mosaic systems & obi versions 1.0.15 & obi 1.2.05 together with a 3rd party treatment planning system like adac pinnacle or cms xio can cause patient positioning errors.

Device

  • Model / Serial
    Model Catalog: PRIOR V1.3) (Lot serial: H080112); Model Catalog: H08 (Lot serial: H080194); Model Catalog: VERSION 1.0.15 (Lot serial: H080112); Model Catalog: H08 (Lot serial: H080112); Model Catalog: PRIOR V1.3) (Lot serial: H080220); Model Catalog: OBI 1.2.05 (VERSIONS (Lot serial: H080220); Model Catalog: VERSION 1.0.15 (Lot serial: H080220); Model Catalog: H08 (Lot serial: H080220); Model Catalog: PRIOR V1.3) (Lot serial: H080059); Model Catalog: OBI 1.2.05 (VERSIONS (Lot serial: H080059); Model Catalog: VERSION 1.0.15 (Lot serial: H080059); Model Catalog: H08 (Lot serial: H080059); Model Catalog: PRIOR V1.3) (Lot serial: H080219); Model Catalog: OBI 1.2.05 (VERSIONS (Lot serial: H080219); Model Catalog: VERSION 1.0.15 (Lot serial: H080219); Model Catalog: H08 (Lot serial: H080219); Model Catalog: PRIOR V1.3) (Lot serial: H080060); Model Catalog: OBI 1.2.05 (VERSIONS (Lot serial: H080060); Model Catalog: VERSION 1.0.15 (Lot serial: H080060); Model Catalog: H08 (Lot serial: H080060); Model Catalog: PRIOR
  • Product Description
    On-Board Imager
  • Manufacturer

Manufacturer