Recall of OMEGA PLUS STANDARD LAG SCREW

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    29279
  • Event Risk Class
    II
  • Event Initiated Date
    2017-04-17
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Stryker received a complaint regarding a standard lag screw omega 110mm length. it was reported that the lag screw was not completely cannulated and the guide wire could not go through the cannula of the screw. no harm reported. the investigation revealed that the obstruction is related to a metal burr in the cannula of the screw which was not removed during manufacturing. the non-conformance is limited to 2 batches.

Device

  • Model / Serial
    Model Catalog: 3362-5-110 (Lot serial: V06279); Model Catalog: 3362-5-105 (Lot serial: V06269); Model Catalog: 3362-5-110 (Lot serial: V06269); Model Catalog: 3362-5-105 (Lot serial: V06279)
  • Product Description
    OMEGA PLUS STANDARD LAG SCREW
  • Manufacturer

Manufacturer

  • Manufacturer Address
    HAMILTON
  • Manufacturer Parent Company (2017)
  • Source
    HC