Recall of OLYMPUS AU IMMUNOCHEMISTRY SYSTEM - APOLIPOPROTEIN B

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by OLYMPUS AMERICA INC. (DIAGNOSTIC SYSTEMS GROUP).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55083
  • Event Risk Class
    III
  • Event Initiated Date
    2003-08-29
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A review showed there was an increase in control sera recovery of >10% between apo b calibrator internal lot 038 and all siblings of calibrator internal lot 040. verification done on fresh not frozen samples.

Device

Manufacturer

  • Manufacturer Address
    MELVILLE
  • Source
    HC