Recall of NX8200

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by AGFA HEALTHCARE INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    111657
  • Event Risk Class
    II
  • Event Initiated Date
    2010-06-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Attempt to reposition the monitor by pulling or pushing the monitor housing will cause strain on the monitor. depending on the extent of stress and after an extended period of time the monitor may fail at or near the mounting bracket pins.

Device

  • Model / Serial
    Model Catalog: X.0.8200 (Lot serial: ); Model Catalog: E5LHV (Lot serial: ); Model Catalog: E5LJX (Lot serial: )
  • Product Description
    NX8200 workstations used in combination with 19'' Barco monitor and mounting stand
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC