Recall of NOVALIS TX HIGH ENERGY LINEAR ACCELERATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by VARIAN MEDICAL SYSTEMS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    146592
  • Event Risk Class
    II
  • Event Initiated Date
    2011-07-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Due to mechanical strain on the coolant line the coolant system gantry swivel joint may leak into or near electrical components within the gantry or gantry stand housing creating an electric shock hazard to personnel working within the gantry fiberglass housing and possibly leading to the failure of the device to operate.

Device

  • Model / Serial
    Model Catalog: (Lot serial: H294674); Model Catalog: (Lot serial: H140836); Model Catalog: (Lot serial: H294551); Model Catalog: (Lot serial: H294364); Model Catalog: (Lot serial: H294317); Model Catalog: (Lot serial: H294838); Model Catalog: (Lot serial: H294383); Model Catalog: (Lot serial: H294675); Model Catalog: (Lot serial: H294766); Model Catalog: (Lot serial: H294683)
  • Product Description
    NOVALIS TX HIGH ENERGY LINEAR ACCELERATOR
  • Manufacturer

Manufacturer