Recall of NEURO III-SV INTRA-OPERATIVE MRI SYSTEM - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by IMRIS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    91092
  • Event Risk Class
    III
  • Event Initiated Date
    2013-03-19
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During servicing a user error occurred that caused a surgical headlight to be drawn into the magnet causing damage to the light & could potentially cause injury to hospital personnel and/or patients. no hospital staff or patients were involved in this incident.

Device

  • Model / Serial
    Model Catalog: NEURO III-SV (Lot serial: >10 CONTACT MFR.)
  • Product Description
    SYSTEM NEURO II AND NEURO III-SV
  • Manufacturer

Manufacturer

  • Manufacturer Address
    WINNIPEG
  • Manufacturer Parent Company (2017)
  • Source
    HC