Recall of N-85 HANDHELD CAPNOGRAPH/PULSE OXIMETER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by COVIDIEN LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    97288
  • Event Risk Class
    II
  • Event Initiated Date
    2011-04-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Oridion medicalthe manufacturer of filterline products has discovered an issue with specific lots of the filterline h set infant/neonatal (part 006324) distributed with nellcor(covidien) n85 monitor kits. the issue concerns the presence of fine plastic strands inside the endo-tracheal tube side of the airway adapter which may become dislodged and inhaled by the patient.

Device

  • Model / Serial
    Model Catalog: N-85 (Lot serial: >THAN 10 PLS CONTACT MFG)
  • Product Description
    FILTERLINE H SET INFANT/NEONATAL IN N-85 MONITOR KIT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MANSFIELD
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.