Recall of MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    34805
  • Event Risk Class
    II
  • Event Initiated Date
    2010-11-15
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A wire in the connector of the generator grid switch supervisor box in each device might become short-circuited between 12v power and ground causing a system breakdown.

Device

  • Model / Serial
    Model Catalog: 0708MD02 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 722043* (Lot serial: C@21321); Model Catalog: 722044* (Lot serial: C@22934); Model Catalog: 722043* (Lot serial: C@22934); Model Catalog: 722044* (Lot serial: C@21342); Model Catalog: 722043* (Lot serial: C@21342); Model Catalog: 722044* (Lot serial: C@21373); Model Catalog: 722043* (Lot serial: C@21373); Model Catalog: 722044* (Lot serial: C@21321); Model Catalog: 722043* (Lot serial: C@22537); Model Catalog: 722044* (Lot serial: C@22537); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 306); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 307); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 405); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 449); Model Catalog: 722018 (Lot serial: S4F619177A); Model Catalog: ALLURA XPER FD10/10 (Lot serial: 489); Model Catalog: INTEGRIS BH3000 (Lot serial: C@95850); Model Catalog: INTEGRIS H1000 (Lot serial: P20-782); Model Catalog: 722021 (Lot serial: S4F630349A); Model Catalog: INTEGRIS H10
  • Product Description
    MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC