Recall of MONACO 5

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ELEKTA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    107507
  • Event Risk Class
    II
  • Event Initiated Date
    2016-05-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    When dicom exporting a 3d monaco plan and the "composite field sequencing" (cfs) checkbox is selected the y jaws will snap to the port shape when they should remain where they were defined. this occurs when the port shape used to define or edit the mlc positions extends beyond or inside the actual jaw positions.

Device

  • Model / Serial
    Model Catalog: MONACO RTP SYSTEM (Lot serial: VERSIONS 5.10.00 AND HIGHER)
  • Product Description
    MONACO 5
  • Manufacturer

Manufacturer

  • Manufacturer Address
    ATLANTA
  • Manufacturer Parent Company (2017)
  • Source
    HC