Recall of MONACO

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ELEKTA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18563
  • Event Risk Class
    I
  • Event Initiated Date
    2013-07-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    When sample couch parameter files are edited the densities altered and "save" is selected the sample files will be updated with these new user defined densities. subsequently when a software upgrade is loaded under the conditions described below the modified couch densities in these sample files will be overwritten with the default value of 1.000.

Device

  • Model / Serial
    Model Catalog: MONACO (Lot serial: VERSIONS 3.10.00 AND LATER)
  • Product Classification
  • Product Description
    MONACO RTP SYSTEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    ATLANTA
  • Manufacturer Parent Company (2017)
  • Source
    HC