Recall of MOBILETT XP MOBILE X-RAY GENERATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS CANADA LIMITED - HEALTHCARE.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19605
  • Event Risk Class
    III
  • Event Initiated Date
    2007-02-28
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The mounting bolts that secure the single tank fork assembly and the mounting bolts that secure the collimator flange could potentially become loose.

Device

  • Model / Serial
    Model Catalog: 018 18 454 (Lot serial: ); Model Catalog: 018 18 363 (Lot serial: ); Model Catalog: 018 18 447 (Lot serial: )
  • Product Description
    MOBILETT XP
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC