Recall of MOBILEDIAGNOST WDR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    41348
  • Event Risk Class
    II
  • Event Initiated Date
    2014-04-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that under very specific error conditions the system may execute an unintended exposure: - the hand switch is erroneously stuck in pressed position while the system is switched off - the system is powered up with the hand switch in that erroneous condition - the user logs in selects a patient for examination gets a green ready light in the examination context if the tube arm is then moved out of the zero position the system is doing an exposure.

Device

  • Model / Serial
    Model Catalog: 9890 010 8952X (Lot serial: ALL)
  • Product Description
    MOBILEDIAGNOST WDR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source