Recall of MICROSCAN POSITIVE MIC PANELS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    81708
  • Event Risk Class
    III
  • Event Initiated Date
    2017-04-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Beckman coulter did not inform the customer that tigecycline minimum inhibitory concentration (mic) is being transmitted to the lis interface data stream. in order to support communication with all lis systems the labpro software is transmitting the tigecycline mic value. some lis systems can only accept numerical values for antimicrobial agent test results and any interpretive criteria applied to these numerical values are generated by the lis system.

Device

  • Model / Serial
    Model Catalog: B1017-216 (Lot serial: NOT LOT SPECIFIC); Model Catalog: B1017-217 (Lot serial: NOT LOT SPECIFIC)
  • Product Description
    MICROSCAN POS MIC PANEL TYPE 34;MICROSCAN POS COMBO PANEL TYPE 43
  • Manufacturer

Manufacturer