Recall of MICROSCAN MICROSTREP PANEL

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    59312
  • Event Risk Class
    II
  • Event Initiated Date
    2013-03-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    There is an issue of false skipped wells and false susceptible misreads with s. agalactiae affecting multiple antimicrobial agents on microscan msp1 msp5 mf5j and mf7j panels processed in a walkaway system. the effect is the potential for false susceptible results for s. agalactiae if the walkaway system reads are not verified. this issue is not panel lot specific nor is it related to a particular model of walkaway instrumentation. no other streptococcal species are affected.

Device

  • Model / Serial
    Model Catalog: 10444757 (Lot serial: ALL); Model Catalog: B1027-201 (Lot serial: ALL)
  • Product Description
    MICROSCAN MICROSTREP PANEL
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC