Recall of MICROPLATE READER INFINITE 200 PRO

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by TECAN AUSTRIA GMBH.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    52185
  • Event Risk Class
    II
  • Event Initiated Date
    2015-05-07
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    During an internal project tecan has determined that due to a firmware bug incorrect luminescence measurement values are displayed if the actual luminescence measurement value equals 4096 rlu or multiples thereof. it was determined that these incorrect measurement values may potentially impact the application result.

Device

  • Model / Serial
    Model Catalog: 30050303 (Lot serial: >10 CONTACT MFR); Model Catalog: 30019337 (Lot serial: >10 CONTACT MFR); Model Catalog: 30034301 (Lot serial: >10 CONTACT MFR); Model Catalog: 30061442 (Lot serial: >10 CONTACT MFR); Model Catalog: 30016056 (Lot serial: >10 CONTACT MFR); Model Catalog: 30063849 (Lot serial: >10 CONTACT MFR)
  • Product Description
    MICROPLATE READER INFINITE 200 PRO
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GRODIG
  • Manufacturer Parent Company (2017)
  • Source
    HC