Recall of MEDLEY PATIENT CARE SYSTEM - PUMP

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CARDINAL HEALTH CANADA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    40966
  • Event Risk Class
    II
  • Event Initiated Date
    2007-11-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Cardinal health recieved reports of inaccurate flow rate with the alaris pump module related to misassembled (missing bent or broken) springs during manufacturing or servicing of the mechanism assembly.

Device

  • Model / Serial
    Model Catalog: 8100LVP (Lot serial: UNITS DISTRIBUTED BEFORE SEPT.); Model Catalog: 8100LVP (Lot serial: 27 2007.)
  • Product Description
    MEDLEY PUMP MODULE
  • Manufacturer

Manufacturer

  • Manufacturer Address
    VAUGHAN
  • Manufacturer Parent Company (2017)
  • Source
    HC