Recall of MAMMOMAT INSPIRATION SYSTEM - BIOPSY UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    92400
  • Event Risk Class
    II
  • Event Initiated Date
    2014-09-19
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The stereo biopsy devices (serial numbers 2000 through 2090) for mammomat inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. the pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. the needle positioning device may move even with the safety switch being set sideward. if this occurs with the needle being already inserted it could lead to an unintentional patient injury.

Device

  • Model / Serial
    Model Catalog: 10501331 (Lot serial: 10140000-2042); Model Catalog: 10501331 (Lot serial: 10140000-1146)
  • Product Description
    Mammomat Inspiration mammography system w Stereostatic/ Biopsy device