Recall of M3 MICRO-MULTILEAF COLLIMATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BRAINLAB AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    81455
  • Event Risk Class
    II
  • Event Initiated Date
    2009-03-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Potential risk of interference when using brainlab m3 on a varian linear accelerator with the internal varian multileaf collimator not in park mode during treatment: the internal varian f-collimator could interfeere with the intended conformal fields of the brainlab m3 during patient treatment. this may cause a critical under-dosage of the target region.

Device

  • Model / Serial
    Model Catalog: 41600-3 (Lot serial: 01-006-41600); Model Catalog: 41600-6 (Lot serial: 01-006-41600); Model Catalog: 41600-5 (Lot serial: 01-006-41600); Model Catalog: 41600-3 (Lot serial: E41600-3/30793); Model Catalog: 41600-6 (Lot serial: E41600-3/30793); Model Catalog: 41600-5 (Lot serial: 01-044-41600); Model Catalog: 41600-3 (Lot serial: 41600-052-99); Model Catalog: 41600-6 (Lot serial: 41600-052-99); Model Catalog: 41600-5 (Lot serial: 41600-052-99); Model Catalog: 41600-3 (Lot serial: 01-044-41600); Model Catalog: 41600-6 (Lot serial: 01-044-41600); Model Catalog: 41600-5 (Lot serial: E41600-3/30793)
  • Product Description
    M3 MICRO-MLC FOR VARIAN WITH MLC
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MUNICH
  • Manufacturer Parent Company (2017)
  • Source
    HC