Recall of LIN-X LINEARITY CONTROL

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    108465
  • Event Risk Class
    III
  • Event Initiated Date
    2011-02-16
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Beckman coulter inc. identified a problem with the barcode labels for the lin-x (lot numbers 2681530k 2661580k). the unicel dxht 800 coulter cellular analysis system generated the error message 'xml parsing error' when attempting to read the label for any lin-x level prior to analysis. an additional character that was included in the label caused the error.

Device

  • Model / Serial
    Model Catalog: 628029 (Lot serial: 2661580K); Model Catalog: 628029 (Lot serial: 2681530K)
  • Product Description
    LIN-X LINEARITY CONTROL
  • Manufacturer

Manufacturer