Recall of INTOUCH CRITICAL CARE BED

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    86383
  • Event Risk Class
    II
  • Event Initiated Date
    2014-07-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Stryker received complaints which allege that a caster has broken on the top of the caster stem where the brake rod inserts into the caster. if more than one caster on a unit is damaged brake functions may be affected and result in reduced brake holding force. if only one caster on the bed is damaged brake functionality continues to meet specifications.

Device

  • Model / Serial
    Model Catalog: 2131 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 2140 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 2130 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 2141 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    INTOUCH BED CASTER STEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    HAMILTON
  • Manufacturer Parent Company (2017)
  • Source
    HC