Recall of INTERNAL PADDLES OPTION FOR DEFIBRILLATOR / MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    142749
  • Event Risk Class
    II
  • Event Initiated Date
    2010-02-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The discharge switch on the handle in affected devices may stick or fail to actuate and prevent the user from delivering defibrillation therapy. this was caused by a manufacturing defect in the paddle switch assembly. this only affects units with date codes (mmyy) of 0908 1008 1108 1208 (units manufactured between september 1 2008 and december 30 2008).

Device

  • Model / Serial
    Model Catalog: M4742A (Lot serial: 1108); Model Catalog: M4741A (Lot serial: 1108); Model Catalog: M4744A (Lot serial: 1008); Model Catalog: M4743A (Lot serial: 1008); Model Catalog: M4742A (Lot serial: 1008); Model Catalog: M4741A (Lot serial: 1008); Model Catalog: M4744A (Lot serial: 0908); Model Catalog: M4743A (Lot serial: 0908); Model Catalog: M4742A (Lot serial: 0908); Model Catalog: M4741A (Lot serial: 0908); Model Catalog: M4743A (Lot serial: 1108); Model Catalog: M4744A (Lot serial: 1108); Model Catalog: M4741A (Lot serial: 1208); Model Catalog: M4742A (Lot serial: 1208); Model Catalog: M4743A (Lot serial: 1208); Model Catalog: M4744A (Lot serial: 1208)
  • Product Description
    INTERNAL DEFIBRILLATION PADDLES
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC