Recall of INTELLIVUE PATIENT MONITOR MP5T

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22204
  • Event Risk Class
    II
  • Event Initiated Date
    2014-09-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that in time-synchronized automatic/sequence mode the nbp automatic measurement series is stopped inadvertently if the repetition interval is increased manually by the user (e.G. from 5 to 15 minutes) in between two consecutive measurements: - a prompt message "any ongoing nbp measurement and automatic cycle stopped" is displayed and a single tone sounds when the measurement series is stopped. - after the automated measurement series has been stopped nbp numeric repetition time and time to next measurement are no longer displayed i.E. the nbp numeric field is blank.

Device

  • Model / Serial
    Model Catalog: M8105AT (Lot serial: DE21008390); Model Catalog: M8105AT (Lot serial: DE21008383); Model Catalog: M8105A (Lot serial: >10 NUMBERS CONTACT MFG.); Model Catalog: M8105AS (Lot serial: DE47004280)
  • Product Description
    INTELLIVUE PATIENT MONITOR MP5T
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC