Recall of INTELLIVUE MP20 PATIENT MONITOR SYSTEM - MULTI MEASUREMENT SERVER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28437
  • Event Risk Class
    II
  • Event Initiated Date
    2017-08-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    St-segments in a 12-lead ecg may become distorted and the st-value may be inaccurate when an affected monitor automatically activates ecg filtering in the event that a "ecg check cable" or "ecg noisy elec xx" inop condition is triggered and displayed.

Device

  • Model / Serial
    Model Catalog: M3001A (Lot serial: >100 numbers contact mfg.); Model Catalog: M8102A (Lot serial: >100 numbers contact mfg.); Model Catalog: M8105A (Lot serial: >100 numbers contact mfg.); Model Catalog: M3002A (Lot serial: >100 numbers contact mfg.)
  • Product Description
    INTELLIVUE PATIENT MONITOR SYSTEM - MULTI MEASUREMENT SERVER;INTELLIVUE MP2 PATIENT MONITOR - MAIN UNIT;INTELLIVUE MP5 PATIENT MONITOR SYSTEM - MAIN UNIT;INTELLIVUE X2 MULTI-MEASUREMENT MODULE - MAIN UNIT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC