Recall of INTELLIVUE - INFORMATION CENTER IX - SOFTWARE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    80800
  • Event Risk Class
    II
  • Event Initiated Date
    2013-03-06
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has become aware that if a customer creates customized trend scales in the trend review tile and the ix or primary server reboots for any reason the ix(s) will enter a reboot loop. if one ix reboots the issue will impact only those patients monitored on that device. if the primary server reboots all patients across all ix devices will be impacted.

Device

  • Model / Serial
    Model Catalog: M3290B (Lot serial: >10 NUMBERS CONTACT MANUF.)
  • Product Description
    INTELLIVUE-INFORMATION CENTER IX-SOFTWARE
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source