Recall of INSTATRAK 3500 PLUS SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    44159
  • Event Risk Class
    II
  • Event Initiated Date
    2012-11-27
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The reason the recall was intiated was because several potential safety issues have been identified with the use of the device. the navigation system may lock up or restart unexpectedly if two scans from the same modality and same patient are transferred to the system they may combine into one scan transmitter arm movement after registration may result in inaccuracy and users may experience difficulty calibrating the device.