Recall of INNOVANCE D-DIMER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    107457
  • Event Risk Class
    II
  • Event Initiated Date
    2017-01-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    An extrapolation in the innovance d-dimer ous setting can lead to false negative (as well as to false positive) results on the sysmex cs-systems. the issue may only occur under the following conditions: - the initial result is falsely flagged with an antigen excess error that triggers an automatic remeasurement in a 1/19 dilution. and - the raw signal of the remeasurement is below the limit of blank (lob) of the method. under such conditions the low remeasurement raw value may be strongly affected by the signal noise of the optical system and can generate an erroneous low reading. the raw value will be corrected by the dilution factor and can be deduced from calibration curve in a false low range.

Device

  • Model / Serial
    Model Catalog: OPBP07 (Lot serial: 10445980); Model Catalog: OPBP03 (Lot serial: 10445979); Model Catalog: OPBP07 (Lot serial: 10445979); Model Catalog: OPBP03 (Lot serial: 10445980)
  • Product Description
    Innovance d-dimer
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC