Recall of INGENUITY CT SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    133689
  • Event Risk Class
    III
  • Event Initiated Date
    2016-11-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Philips has received a complaint indicating that the non-gated bolus tracking clinical scan failed to initialize with the system displaying the error message "a problem has occurred- please retry. if the problem persists contact service." this resulted in the study being cancelled with a loss of bolus. if this were to re-occur it may necessitate a rescan.

Device

  • Model / Serial
    Model Catalog: 728326 (Lot serial: 320227); Model Catalog: 728323 (Lot serial: 320227); Model Catalog: 728326 (Lot serial: 30003); Model Catalog: 728323 (Lot serial: 30003); Model Catalog: 728326 (Lot serial: 30004); Model Catalog: 728323 (Lot serial: 30004); Model Catalog: 728326 (Lot serial: 32021); Model Catalog: 728323 (Lot serial: 321123); Model Catalog: 728326 (Lot serial: 321123); Model Catalog: 728323 (Lot serial: 32021); Model Catalog: 728231 (Lot serial: 9074); Model Catalog: 728231 (Lot serial: 10213); Model Catalog: 728231 (Lot serial: 9920); Model Catalog: 728231 (Lot serial: 9923)
  • Product Description
    INGENUITY CT SYSTEM;BRILLIANCE 40/64 SLICE CT SYSTEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC