Recall of INGENUITY CT SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by PHILIPS ELECTRONICS LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    99982
  • Event Risk Class
    III
  • Event Initiated Date
    2015-07-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Some scans unexpectedly set the axial (z axis) scan length to a greater length than the user set in the plan. the issue occurs only when the following 3 conditions are met: -the plan employs dual-surview -the plan requests either a sagittal or a coronal reconstruction -the gantry is tilted.

Device

  • Model / Serial
    Model Catalog: 728326 (Lot serial: 320123); Model Catalog: 728323 (Lot serial: 320123); Model Catalog: 728323 (Lot serial: 30003); Model Catalog: 728326 (Lot serial: 30003); Model Catalog: 728323 (Lot serial: 30004-1); Model Catalog: 728326 (Lot serial: 30004-1); Model Catalog: 728326 (Lot serial: 32021); Model Catalog: 728323 (Lot serial: 32021); Model Catalog: 728326 (Lot serial: 320227); Model Catalog: 728323 (Lot serial: 320227); Model Catalog: 4550 110 04011 (Lot serial: 9920); Model Catalog: 4550 110 04011 (Lot serial: 9923)
  • Product Description
    INGENUITY CT SYSTEM