Recall of INFUSION PUMPS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SMITHS MEDICAL CANADA LTD. (SIMS).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    82198
  • Event Risk Class
    II
  • Event Initiated Date
    2006-05-04
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    When device is exposed to direct and intense light the light penetrates the housing and impacts the optical sensor resulting in a "motor not running" or "motor rate error" causing the pump to stop infusing.

Device

  • Model / Serial
    Model Catalog: 3010 (Lot serial: S/N M02387 TO M03093); Model Catalog: 3010A (Lot serial: S/N M02408 TO M22307); Model Catalog: 3500 (Lot serial: S/N M02408 TO M22307)
  • Product Description
    SEE 33-01 - TOO LONG