Recall of INFUSION PUMP PATIENT CONTROLLED AMBULATORY PCA II WITH CARTRIDGE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    33501
  • Event Risk Class
    II
  • Event Initiated Date
    2006-01-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Manufacturer has received reports of delivery of unrequested pca due to an electrical short in the pca circuit simulating repeated pressing of the pca button.

Device

  • Model / Serial
    Model Catalog: 2L3104R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2L3104 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2L3107R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2L3107 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2L3105 (Lot serial: ALL SERIAL NUMBERS)
  • Product Description
    INFUSION PUMPPATIENT CONTROLLED AMBULAT
  • Manufacturer

Manufacturer