Recall of INFINITY DELTA MODULAR PATIENT MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DRAEGER MEDICAL CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    88502
  • Event Risk Class
    II
  • Event Initiated Date
    2010-06-04
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Investigation determined that the monitor's fixed keys may become inoperative or activate without user interaction due to a contamination present in the fixed key panel material. this is a concern because the worst case scenario of this malfunction could be that the monitor may discharge to a patient automatically.

Device

  • Model / Serial
    Model Catalog: MS18597 (Lot serial: MORE THAN 10 SERIAL NUMBERS); Model Catalog: MS18597 (Lot serial: SEE MANUFACTURER.); Model Catalog: MS18852 (Lot serial: 5399664949); Model Catalog: MS18852 (Lot serial: 6000446771); Model Catalog: MS18852 (Lot serial: 5399533850); Model Catalog: MS18852 (Lot serial: 5399532753); Model Catalog: MS18852 (Lot serial: 5399532156); Model Catalog: MS18852 (Lot serial: 5399498745); Model Catalog: MS18852 (Lot serial: 5399494054); Model Catalog: MS18852 (Lot serial: 5399469251); Model Catalog: MS18852 (Lot serial: 5399468849); Model Catalog: MS18852 (Lot serial: 5399463952); Model Catalog: MS18852 (Lot serial: 5399662354); Model Catalog: MS18852 (Lot serial: 6000796170)
  • Product Description
    DELTA MONITOR
  • Manufacturer

Manufacturer