Recall of INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE -

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by DRAEGER MEDICAL CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    119844
  • Event Risk Class
    II
  • Event Initiated Date
    2014-02-19
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The battery capacity of the optional ps500 of the infinity acs workstation critical care (used in connection with evita infinity v500 babylog vn500 or evita v300) will not last as expected. in some cases neither the "battery low" nor the "battery depleted" alarm was given when the remaining battery capacity fell below 10%. however when the battery depleted toatlly the power fail alarm was generated.

Device

  • Model / Serial
    Model Catalog: 8417400 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    INFINITY ACUTE CARE SYSTEM WORKSTATION
  • Manufacturer